Presource kits and trays, labeled as: 1) Cardiovascular Procedure Kit, Catalog Number SAN29ANCC3; 2) Cardiovascular Procedure Kit, Catalog Number SAN29ANCC4; 3) Cardiovascular Procedure Kit, Catalog Number SAN29ANCCL; 4) Cardiovascular Procedure Kit, Catalog Number SAN29ANCCM; 5) Cardiovascular Procedure Kit, Catalog Number SAN33CDCAS; 6) Cardiovascular Procedure Kit, Catalog Number SAN33CDCAT; 7) Cardiovascular Procedure Kit, Catalog Number SANOCAENSA; 8) Cardiovascular Procedure Kit, Catalog Number SCV29ANCCA; 9) Left Heart Kit, Catalog Number SAN73LHKCC; 10) Left Heart Kit, Catalog Number SAN73LHKC6
FDA Recall Z-2329-2024 · Posted 2024-07-10
Advertisement
Recalling FirmCardinal Health 200, LLC
Recall StatusOpen, Classified
Product CodeOEZ
Date Initiated2024-05-29
Product Quantity17445 units
Reason for Recall
Cardinal Health is following the FDA's recommendations regarding the disposition of affected Jiangsu Shenli Medical Production Co. Ltd (Shenli) Syringes in the Merit Medical supplied finished kits. Merit Medical supplied finished kits are included as components in Presource kits. According to the FDA s communication, these Shenli syringes, located inside a Merit Medical kit component, cannot be distributed for use as they could contain a significant change or modification in design which could significantly alter the safety or effectiveness of the finished kit product, resulting in the risk of patient harm such as inaccurate dosing and a leaking device.
Firm Action
Cardinal Health issued an URGENT: PRODUCT RECALL notice to its consignees on 05/28/2024 via letter. The notice explained the issue, potential risks involved, and requested the following:
"If you locate affected product within your facility, please complete the following:
1. QUARANTINE affected kits & ALERT your clinicians of this action
2. AFFIX a WARNING LABEL to the front of each Presource kit so that it is clearly visible to clinicians, instructing them to remove and discard all syringes within the affected Merit Medical kit in the Presource product.
3. RETURN a copy of the enclosed Acknowledgment Form confirming your receipt of this action and the completion of labeling product via fax to 614.652.9648.
Additional labels will be sent to your facility upon request. Please send an email to:
[email protected]
For component credit or replacement, please contact the following:
" Hospitals Cardinal Customer Service at 866.551.0533
" Distributors Distributor Management at 800.635.6021
" Federal Government facilities Government Customer Service at 800.444.1166
If you are a Distributor and have further distributed affected kits, please notify your customers of this recall."
Expansion letter was sent via FedEx Express on 6/20/2024
Root Cause
Device Design
Distribution Pattern
Nationwide
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.