Mazor X Surgical System Positioner Type II, REF: ASM0214-02

FDA Recall Z-2330-2020 · Posted 2020-06-09

Recall StatusOpen, Classified
Product CodeOLO
Date Initiated2019-12-05
Product Quantity177 Systems

Reason for Recall

There is a potential for the surgical system to detach from the operational room table unexpectedly. This could result in the surgical system falling onto the patient resulting in blunt injury [hemorrhage, hematoma, bleeding, or fracture(s)]

Firm Action

The firm originally sent a "Medical Device Correction " letter to affected consignees on December 5, 2019. On November 9, 2020, a "Update - Medical Device Correction" letter was sent out to all affected customers. The purpose of the letter was to update consignees that since the initiation of the recall (12/5/2019), Medtronic had developed a fix to permanently correct the issue associated with the Mazor X Surgical system. The notice further stated that upon availability of parts, Medtronic will schedule a service visit to correct the system at no charge. Medtronic expects to initiate these service visits within the next six months. Pending completion of service to correct the Positioner Type II system, continue to follow the modified mounting instructions. The firm requested that consignees do the following: 1. Please sign and date the Customer Confirmation Form enclosed with this letter to confirm that you have reviewed the information included in this notification with all users of the affected Mazor X" Surgical System, including physician users. 2. Return the form to Medtronic via email at [email protected] or via fax at 651-367-7075 within 30 days of receipt. 3. If you have any questions regarding this notification, please contact Medtronic at 1-888-826-5603. Medtronic will initiate service visits to correct systems on-site at customer locations when the kits become available. This is expected to begin by February 2021. View spreadsheet for updated

Root Cause

Device Design

Distribution Pattern

US: ALAR,AZ,CA,CO,CT,FL,GA,IA,ID,IL,IN,KY,MA,MI,MN,MO,NC,ND,NL,NY,OH,PA,SC,TN,TX,UT,VA,WA,WI,WV, OUS: Australia, China, India, Israel, Romania, Singapore, Thailand, United Kingdom, Switzerland, Spain, Germany, Belgium,