Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Open, Classified

smith&nephew R3 ZERO HOLE HA COATED HEMISPHERICAL STIKTITE COATED SHELL, manufactured in the following sizes: Size REF/product number 42MM 71332242 46MM 71332246 48MM 71332248 50MM 71332250 52MM 71332252 54MM 71332254 56MM 71332256 58MM 71332258 60MM 71332260 62MM 71332262 64MM 71332264 Orthopedic implant component.

FDA Recall Z-2332-2020 · Posted None

Recalling FirmSmith & Nephew, Inc.
Recall StatusOpen, Classified
Product CodeMBL
Date Initiated2020-04-22
Product Quantity1450 devices

Reason for Recall

A manufacturing error resulted in out of specification R3 Acetabular Shells

Firm Action

The firm notified its consignees by letter (delivered by Fed Ex) and email beginning on 04/22/2020. Consignees are asked to inspect inventory, locate and quarantine affected devices. Affected product will be returned to Smith & Nephew. Questions or concerns regarding this recall may be directed to: [email protected].

Root Cause

Process control

Distribution Pattern

Worldwide

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑