Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

BD" Mouse IgG2a Isotype Control APC X39 ASR

FDA Recall Z-2333-2021 · Posted None

Recall StatusTerminated
Product CodeMVU
Date Initiated2021-07-22
Date Terminated2023-03-24
Product Quantity23 reagents

Reason for Recall

Complaints have been received for performance issues related to high background and staining. this issues is related to a manufacturing process error that can potentially cause a false positive staining due to contamination with another antibody.

Firm Action

On 07/22/2021, the firm sent an "URGENT MEDICAL DEVICE RECALL" Notification letter via FedEx to customers to inform them that the Recalling Firm received 3 complaints for performance issues related to high background and staining. Our investigation has confirmed there was an error during the manufacturing process that can potentially cause a false positive staining due to contamination with another antibody. Customers are instructed to: 1) Immediately review their inventory for the specific Catalog and lot number listed in the Recall Notification above. Destroy all product subject to the recall following their institution process for destruction. 2) Share this recall notification with all users of the product within their facility to ensure that they are also aware of this recall. If the affected product was used to generate test results, review potential impact to test results following their institution's policies. 3) Complete the attached Customer Response Form and return to the BD contact noted on the form. Even if you do not have any of the affected lots in your inventory, please complete the Customer Recall Response Form indicating you have zero (0) quantity and return as indicated. Actions being taken by the Recalling Firm: 1) An investigation has been initiated to identify root cause and corrective actions to prevent recurrence. 2) The Recalling Firm will provide replacement for any discarded product. For questions or additional assistance contact North American Regional Complaint Center at 1-844-8BD-LIFE (1-844-823-5433) M-F 8am -5pm CT

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide distribution - U.S. Nationwide distribution in the states of CT, FL, KY, NY, and PA. The country of Taiwan.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑