smith&nephew R3 MULTI HOLE HA COATED HEMISPHERICAL STIKTITE COATED SHELL, manufactured in the following sizes: Size REF/product number 48MM 71338733 50MM 71338734 52MM 71338735 54MM 71338736 56MM 71338737 58MM 71338738 62MM 71338740 64MM 71338741 Orthopedic implant component.
FDA Recall Z-2336-2020 · Posted None
Advertisement
Recalling FirmSmith & Nephew, Inc.
Recall StatusOpen, Classified
Product CodeMBL
Date Initiated2020-04-22
Product Quantity58 devices
Reason for Recall
A manufacturing error resulted in out of specification R3 Acetabular Shells
Firm Action
The firm notified its consignees by letter (delivered by Fed Ex) and email beginning on 04/22/2020. Consignees are asked to inspect inventory, locate and quarantine affected devices. Affected product will be returned to Smith & Nephew.
Questions or concerns regarding this recall may be directed to: [email protected].
Root Cause
Process control
Distribution Pattern
Worldwide
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.
Advertisement