Estrone RIA Catalog Number DSL8700 UDI Code: 15099590211615 Estrone RIA is an in vitro diagnostic manual medical device used by healthcare professionals for the quantitative measurement of estrone in human serum and plasma.

FDA Recall Z-2336-2025 · Posted 2025-08-18

Recalling FirmImmunotech A.S.
Recall StatusOpen, Classified
Product CodeCGF
Date Initiated2025-06-12
Product Quantity279 units

Reason for Recall

Estrone RIA kit lots may rarely give falsely high patient results which might cause Temporary injury of Remote probability or rarely give falsely low patient results which might cause Permanent injury of Remote probability. Only in some (very rare) cases may the issue occur.

Firm Action

On 06/12/2025, the firm sent an "URGENT MEDICAL DEVICE RECALL" Letter to customers informing them that Beckman Coulter determined that the Estrone RIA kit might be affected by an isolated coated tubes inhomogeneity issue causing falsely elevated results or falsely decreased results. Customers are instructed to: Beckman Coulter recommends sharing the content of the Recall Letter with their laboratory and/or medical director regarding the need to review previous patient test results. Note- All affected product have expired (expiry dates 12/31/2024 thru 04/22/2025) For questions, contact Customer Support Center: Website: http://www.beckmancoulter.com Email: [email protected]

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide - U.S. Nationwide distribution in the states of NC and the countries of Argentina, Austria, Belgium, Canada, France, Germany, Greece, Hungary, Italy, and Netherlands.