Angio-Seal Evolution Vascular Closure Device, Cat. No. C610135
FDA Recall Z-2337-2018 · Posted None
Advertisement
Recalling FirmTerumo Medical Corp
Recall StatusTerminated
Product CodeMGB
Date Initiated2018-04-09
Date Terminated2019-04-19
Product Quantity240
Reason for Recall
Two lots of product were released for distribution prior to the completion of all required validation and lot acceptance activities. Although sterile, products from these lots may not function as intended and therefore should not be used.
Firm Action
On April 9, 2018, the firm sent Urgent Voluntary Medical Device Recall letters and Inventory Confirmation Forms to customers via FedEx. The letter described the product issue.
Customers were asked to:
1. Review your Angio-Seal Evolution inventory immediately to identify and isolate affected inventory in order to prevent future use.
2. Your Terumo Territory Manager will visit your facility to confirm product inventory and assist you in completing the enclosed Recall Acknowledgement/
Inventory Confirmation Form. This required documentation, along with the enclosed Fed-Ex product shipment labels will facilitate your return of affected Angio-Seal Evolution product to Terumo Medical Corporation.
Root Cause
Release of Material/Component prior to receiving test results
Distribution Pattern
Distributed to accounts in AL, CA, FL, IA, IN, MI, MO, NC, NY, SD, TN, and TX.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.