Terminated

ARTHROSCOPY PACK - (1) MAYO STAND COVER REINFORCED LIF (1 0) GAUZE SPONGES 4" X 4" 16PL Y (1) TUBE SUCTION CONNECTING X" X 12' LIF (1) IMP. STOCKINETTE (1) SURGICAL BLADE #11 STAINLESS STEEL (1) ANESTHESIA SPINAL NEEDLE 18G X 3 ~ (1) SYRINGE Sec WITH NEEDLE 21G X 1 ~ LUER LOCK LIF (2) ELASTIC BANDAGE 6" X 5yrd LIF (1) TUR Y SET 81 "2.1m REGULAR CLAM (2) NEEDLE HYPODERMIC 18G X 1 ~ (1) TABLE COVER 50" X 90" REINFORCED LIF (5) TOWEL ABSORBENT 15" X 20" LIF (2) DRAPE UTILITY WITH TAPE LIF (1) DURAPREP SURGICAL SOLUTION 26m I LIF (1) U-DRAPE 60" X 70" WITH TAPE SPLIT (1) Pr. NEUTRALON SURGEON GLOVES #7.5 (1) Pr. NEUTRALON SURGEON GLOVES #8 (1) Pr. NEUTRALON SURGEON GLOVES #8.5 (2) GOWN IMPERVIOUS REINFORCED X-LARGE SMS (10) GAUZE SPONGES 4" X 4" 16PLY XRD (2) STRIPS TAPE (1) SHEET ARTHROSCOPY T STD SMS WITH POUCH LIF (1) DRAPE VIDEO CAMERA LIF EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents.

FDA Recall Z-2338-2014 · Posted 2014-08-29

Recalling FirmCustomed, Inc
Recall StatusTerminated
Product CodeFWZ
Date Initiated2014-05-20
Date Terminated2017-08-14
Product Quantity28 lots; 2620 units (multiple units per lot)

Reason for Recall

Customed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.

Firm Action

Customed Inc. sent a Recall Notification letter dated June 3, 2014 via email and followed by letter to all affected customers. The letter identified the affected products, problem and actions to be taken. The letter informed customers if the affected products are in their inventory to: 1. Remove products from sale (quarantine them) and cease distribution. 2. Complete and return the enclosed acknowledgement form to Customed Inc .Recall Coordinator. Affected products should be returned to Customed Inc for credit. Customers with questions were instructed to contact Customed Inc. Recall Coordinator at 1-787-801-0100 Ext. 7540

Root Cause

Package design/selection

Distribution Pattern

Worldwide Distribution - US (nationwide) in the state of Florida including Puerto Rico (USA); & USVI (St John) and the countries of Tortola, Trinidad, Dominican Republic, Costa Rica and Mexico.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑