Cannulated Driver Handle, Catalog #: 210-00-003, Rx Only, Non Sterile

FDA Recall Z-2338-2018 · Posted None

Recall StatusTerminated
Product CodeMDM
Date Initiated2018-03-27
Date Terminated2020-06-04
Product Quantity94

Reason for Recall

Firm is voluntarily recalling cannulated driver handles (210-00-003 and 210-00-004) that may demonstrate a degradation of function with the adapter collar sticking, and/or not retaining the mating adapter male shaft.

Firm Action

On 03/27/18, the firm sent Medical Device Recall notices to customers informing them to: 1) Check inventory stock for identified devices; 2) Locate and remove and/or discontinue use of identified devices; 3) Return all related product(s) to the firm in accordance with an included return response / acknowledgement form. Customers were instructed to contact the firm at 1-800-495-2919 to organize product returns.

Root Cause

Other

Distribution Pattern

U.S. States: NE, MA, GA, TX, OR, AZ, FL, OH, WA, NJ, UT, VA, CA, MN, PA, DE, OK, KS, WI, MT, IA, MO