Open, Classified

Gore Molding & Occlusion Balloon Catheter, REF Catalogue Number MOB37, SN XXXXXXXX, 10 mm - 37 m 10 Fr, Guidewire Compatibility 0.035", STERILE EO, CE2797, UDI: (00)733132639489

FDA Recall Z-2340-2021 · Posted None

Recall StatusOpen, Classified
Product CodeMJN
Date Initiated2021-07-12
Product Quantity3,364 units

Reason for Recall

Complaints received concerning Balloon Catheter leakage from the guidewire lumen and the y-hub during preparation resulting in either insufficient inflation or an inability to inflate.

Firm Action

On July 23, 2021 issue a "Urgent Medical Device Recall" notification to affected consignees via UPS. In addition to informing consignees about the recall the customer notification ask consignees to take the following actions: 1. Identify and return any unused devices within the scope of this voluntary recall. 2. Please complete and sign the enclosed CUSTOMER RESPONSE FORM and return to [email protected] within 2 weeks of receipt of this notification. 3. Please share this letter with others in your hospital or clinic as appropriate. 4. If a listed device has been used, there is no patient-follow up needed and there are no further actions needed other than informing Gore the device was used. Please indicate the used devices on the CUSTOMER RESPONSE FORM and return to [email protected] within 2 weeks of receipt of this letter. No further action is needed.

Root Cause

Process change control

Distribution Pattern

Worldwide distribution - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, WA, WI, WV, WY and the countries of AUS, CAN, CHE, DEU, DNK, ESP, FIN, FRA, GBR, GRC, HRV, HUN, IRL, ISR, ITA, JPN, MEX, NLD, NOR, POL, PRT, SMR, SWE.