Merge HEMO software.

FDA Recall Z-2341-2016 · Posted 2016-08-04

Recall StatusTerminated
Product CodeDQK
Date Initiated2012-07-02
Date Terminated2017-06-26
Product Quantity37

Reason for Recall

The International Normalized Ratio (INR) value displayed in the study report does not match the value that is imported on the pre-procedure labs screen from HL7

Firm Action

The firm issued an "Important Product Information" bulletin to customers dated 7/2/2012 via e-mail letting them know there was an upgrade available. On 1/30/2016, the recalling firm initiated phone calls to their customers letting them know there was an upgrade available and if they declined it, they would like a written response.

Root Cause

Software design

Distribution Pattern

The medical facilities who currently have the affected versions are located nationwide. There was government distribution but no military or foreign distribution.