Safety Sideport Knife 1.0mm 45, REF 378210 Product Usage: Indicated to make stab incisions during ophthalmic surgery
FDA Recall Z-2342-2018 · Posted None
Advertisement
Recalling FirmBeaver Visitec
Recall StatusTerminated
Product CodeEMF
Date Initiated2018-05-14
Date Terminated2021-06-08
Product Quantity1354 units
Reason for Recall
Blades are loose in handles and may fall out due to non cured epoxy
Firm Action
On May 14, 2018, Beaver-Visitec International (BVI) issued PRODUCT RECALL NOTICES to customers.
Customers were advised to take the following actions:
-Inspect stock, quarantine and destroy product per internal procedures.
-Complete and return the response form via email to [email protected]
- Provide objective evidence of product destruction with BVI so that credit to customer accounts can be applied.
Customers with questions may contact the firm by calling 866-906-8080.
Root Cause
Nonconforming Material/Component
Distribution Pattern
US Nationwide in the states of CA, IA, CO, HI, IA, MD & TX