GE HealthCare OEC 9900 Elite C-arm Systems.
FDA Recall Z-2342-2025 · Posted 2025-08-22
Advertisement
Recalling FirmGE Medical Systems, LLC
Recall StatusOpen, Classified
Product CodeOWB
Date Initiated2025-07-02
Reason for Recall
GE HealthCare has become aware of a non-conformance to 21 CFR 1020.30 (k) in certain OEC 9800 and OEC 9900 Elite C-arm Systems. Specifically, impacted systems are missing the full layer of lead tape in the collimator covers, which will lead to a small increase in radiation exposure to the patient and/or operator. The missing lead tape is not detectable by the user.
Firm Action
GE HealthCare will send a customer letter providing safety instructions as an immediate
mitigation, to communicate the risk of additional radiation and that the equipment should only be used by personnel properly trained and authorized in the use of ionizing radiation. A user should observe proper safety practices during equipment operation as required by the operator manual.
GE HealthCare will correct all affected OEC 9800 and OEC 9900 Elite systems (28 units) by
visually inspecting for the presence of lead shield tape on the collimator cover and replacing the collimator cover if it is found non-compliant, at no cost to the customer.
Root Cause
Radiation Control for Health and Safety Act
Distribution Pattern
US Nationwide distribution.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.