Terminated

MAGNETOM Spectra magnetic resonance diagnostic devices Model # 10837643

FDA Recall Z-2344-2019 · Posted None

Recall StatusTerminated
Product CodeLNH
Date Initiated2019-07-09
Date Terminated2020-04-23
Product Quantity2

Reason for Recall

Potential of an open port in the quench venting system when using the optional horizontal outlet on the quench line assembly.

Firm Action

1. review the System Owner Manual chapter 1.3.3 Quench emergency plan and the section regarding defective vent line. 2. follow the standard emergency procedures you have in place in the event of system failures. 3.A Siemens Service Technician will conduct an inspection to verify that your system has a vertical outlet. If a vertical outlet is present, no further action is required. In the event your system is equipped with a horizontal outlet, the Siemens Service Technician will ensure that the port for the drop-off plate is adequately blanked off or sealed. 4.service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organization for an earlier appointment at 1-800-888-7436. 5. Please review, understand, and confirm acknowledgement of the information and instructions outlined in this Customer Safety Advisory Notice. By signing this document you are acknowledging that you have read and understand the content therein. If you have any question, please contact Meredith Adams, Telephone 610-448-6461 and E-Mail: [email protected]

Root Cause

Device Design

Distribution Pattern