Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

IMPLANTE DE ORTHOPEDIA- 1) GOWN IMPERV. XTRA REINF. XL SMS T/WRAP LEVEL IV (1) CAUTERY TIP POLISHER LIF (1) DURAPREP SURGICAL SOLUTION 26ml LIF (1) BAG SUTURE FLORAL LIF (1) TUBE SUCTION CONNECT. Y.!" X 12' L/F (1) CAUTERY PENCIL ROCKER SWITCH LIF (2) *Pr. SURGEON NEUTRALON GLOVE# 8 LATEX (2) BLADE SURGICAL# 10 STAINLESS STEEL (1) BAG GLASSINE (1) STOCKINETIE IMPERVIOUS 14" X 48" LIF (2) LITE GLOVE LIF (1) SPLIT SHEET W/ADHES 108" X 77" STD SMS UF (1) MAYO STAND COVER REINFORCED LIF (4) DRAPE UTILITY WITH TAPE LIF (5) ABSORBENT TOWELS 15" X 20" LIF (5) LAP SPONGES PRE-WASH XRD LIF (2) DRAPE SHEET 70" X 100" STD SMS LIF (1) MAYO TRAY LARGE (2) * Pr. SURGEON NEUTRALON GLOVE #8% LATEX (1) STAPLE SKIN WIDE 35 ST. LIF (1) TOP DRAPE W/ADHES. 108" X 50" STD SMS LIF (1) BULB SYRINGE 60cc (2) BANDAGE ELASTIC 6" X 5"yds LIF (2) UTILITY BOWL 32oz (1) UTILITY BOWL 16oz LIF (1) SKIN MARKER WITH RULER (1) INCISE DRAPE ANTIMICROBIAL 23" X 17" LIF (2) ABDOMINAL PAD 8" X 7 W (1) U-DRAPE 60" X 70" WIT APE SPLIT 6" X 21' (2) GOWN IMPERVIOUS REINFORCED XL (1) YANKAUER SUCTION TUBE WITHOUT VENT UF (2) COVER TABLE BTC 77" X 86" HD FULL COVERAGE EO sterilized surgical convenient packs packaged into polyester/polyethylene breather bag with Tyvek double vents.

FDA Recall Z-2346-2014 · Posted 2014-08-29

Recalling FirmCustomed, Inc
Recall StatusTerminated
Product CodeOJH
Date Initiated2014-05-20
Date Terminated2017-08-14
Product Quantity16 lots; 1041 units (multiple units per lot)

Reason for Recall

Customed has determined that there is the possibility that packaging integrity may be compromised on different catalog numbers (multiple lots). This could result in an injury to the patient due to product contamination or loss of sterility condition.

Firm Action

Customed Inc. sent a Recall Notification letter dated June 3, 2014 via email and followed by letter to all affected customers. The letter identified the affected products, problem and actions to be taken. The letter informed customers if the affected products are in their inventory to: 1. Remove products from sale (quarantine them) and cease distribution. 2. Complete and return the enclosed acknowledgement form to Customed Inc .Recall Coordinator. Affected products should be returned to Customed Inc for credit. Customers with questions were instructed to contact Customed Inc. Recall Coordinator at 1-787-801-0100 Ext. 7540

Root Cause

Package design/selection

Distribution Pattern

Worldwide Distribution - US (nationwide) in the state of Florida including Puerto Rico (USA); & USVI (St John) and the countries of Tortola, Trinidad, Dominican Republic, Costa Rica and Mexico.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑