Terminated

KyphX Osteo Introducer System with blunt tip introducer stylet (as contained in the KyphoPak Tray, First Fracture, 15/3 Xpander), catalog number KPT1502, manufactured by Medtronic Spine LLC. Device is used to access vertebral body during balloon kyphoplasty, a procedure designed to reduce and stabilize vertebral body compression fractures (VCFS).

FDA Recall Z-2347-2008 · Posted 2008-09-18

Recalling FirmKyphon Inc
Recall StatusTerminated
Product CodeLXH
Date Initiated2008-06-23
Date Terminated2009-10-07
Product Quantity310 units

Reason for Recall

Some products may contain the incorrect introducer stylet. If the treating physician is unfamiliar with the use and design of the diamond tip stylet, there is a possibility of serious injuring occurring with the use of the diamond tip stylet during balloon kyphoplasty procedure. Other risks include delay to surgery and inconvenience to patient.

Firm Action

Recall initiated on June 23, 2008. A recall letter and reconciliation form was sent out via e-mail to sales representatives for hand delivery to US hospital consignees. The consignees were asked to review the letter and complete/return the reconciliation form to the firm. If you have any questions contact Kyphon, Inc. at 1-408-548-6740.

Root Cause

Mixed-up of materials/components

Distribution Pattern

Class II Recall - Nationwide Distribution.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑