Steris Small Renaissance Eagle 3000 -- Remanufactured Gravity Models 3011 and 3021 and Prevacuum Models 3013 and 3023. Products are designed for efficient sterilization of non-porous and porous, heat and moisture-stable materials used in healthcare facilities.

FDA Recall Z-2350-2008 · Posted 2008-09-17

Recalling FirmSteris Corp
Recall StatusTerminated
Product CodeFLE
Date Initiated2008-03-06
Date Terminated2010-09-28
Product Quantity148 units

Reason for Recall

The firm discovered that some Eagle 3000 units may have a door post pin missing which could cause incomplete or no engagement of the radial arms into the lock rails of the door. The sterilizer is designed with switches that prevent cycle initiation if the door is not locked. However, in the event the door switches are not properly adjusted they may falsely indicate that the door is locked even t

Firm Action

On March 5, 2008, Steris issued an Urgent Product Correction. Beginning March 6, 2008 customers were notified by telephone. The firm instructed their customers to immediately discontinue using the small Renaissance Eagle 3000 steam sterilizer(s) until inspected and corrected by a service technician. Steris Service technicians were dispatched to affected customers to inspect and correct the sterilizers as appropriate. If you have any questions contact Holly Wright Lee at (404) 392-7019 or 1-800-548-4873 ext. 27019.

Root Cause

Employee error

Distribution Pattern

Class II Recall - Worldwide Distribution --- USA including states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, IL, IN, KY, LA, MD, MI, MN, MO, MT, NC, NJ, NV, NY, OH, OR, PA, SC, TN, TX, UT, VA, WI, and WY, and countries of Canada and Mexico.