CELL-DYN 3700 CS and SL, Hematology Analyzer manufactured by Abbott Laboratories, Santa Clara, CA. Waste Reservoir Assembly Part number 8921174802.

FDA Recall Z-2351-2010 · Posted 2010-09-02

Recalling FirmAbbott Laboratories
Recall StatusTerminated
Product CodeGKZ
Date Initiated2010-04-16
Date Terminated2010-12-06
Product Quantity69 units total

Reason for Recall

Epoxy holding the two parts of the waste reservoir together fails, potentially resulting in biohazardous fluid leaks.

Firm Action

Product Recall letters were sent on 04/16/2010. The letters identified the affected product, explained the reason for recall, the patient impact/user safety, stated the necessary actions, and contact information. Users are to discontinue use until the Waste Chamber(s) have been replaced and follow the instructions in the Hazards section. In addition, appropriate protective equipment needs to be worn along with following their recommended laboratory procedures dealing with spills. Users are to contact their local Abbott Customer Support rep. US customers should be direct their questions to Customer Support.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Worldwide Distribution -- CA, CT, ID, IN, and MA; and, countries of Australia, Brazil, Canada, China, Colombia, Mexico, Puerto Rico, Singapore, and Uruguay.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →