Accu-Chek SmartView Test Strip, Blood Glucose Monitoring System, Model numbers: 06337538001, 06337546001, 06337562001

FDA Recall Z-2353-2021 · Posted None

Recall StatusOpen, Classified
Product CodeLFR
Date Initiated2021-07-28
Product Quantity1,150,755 (US)

Reason for Recall

Test strip vials may open while inside sealed cartons during shipment. An open vial might expose the test strips to humidity, which might damage the strips and result in inaccurate results, potentially leading to inappropriate therapy decisions and adverse health consequences.

Firm Action

On July 28, 2021, the firm issued an Urgent Medical Device Correction to affected customers regarding the potential for open vials. Customers were warned that open vials may expose the test strips to humidity, which might damage the strips and result in inaccurate results (such as positively biased or falsely too high results). Inappropriate therapy decisions based on inaccurate results could lead to adverse health consequences. ACTIONS REQUIRED BY USERS: 1. Check vials of affected test strips before use. DO NOT use test strips if: - The vial is open or damaged before using the test strips for the first time - The cap is not fully closed - You see any damage to the cap or vial, or - Anything prevents the cap from closing properly 2. DO NOT perform control testing if you open a sealed carton and ay of the vials inside meet the criteria listed above. 3. Contact Accu-Chek Customer Care for product replacement if you open a sealed carton and any of the vials inside meet the criteria listed above by phone at 1-800-858-8072 or on our website at Accu-Chek.com under Contact Us for email or chat. Please have the affected products available. 4. Complete and return the business reply letter to the recalling firm. 5. Dispose of the affected test strips and vial according to your local guidelines. Contact Accu-Chek Customer Care if you have questions regarding the information in this Urgent Medical Device Correction(UMDC) by phone at 1-800-858-8072 or on our website at Accu-Chek.com under Contact Us for email or chat.

Root Cause

Package design/selection

Distribution Pattern

Worldwide distribution - US Nationwide distribution.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑