A LINK Z INTERVERTEBRAL BODY FUSION DEVICE Product Description: A Link Z Ti Unitary Open Interbody 32x24x12 7 A Link Z Ti Unitary Open Interbody 32x24x12 12 A Link Z Ti Unitary Open Interbody 32x24x16 7 A Link Z Ti Unitary Open Interbody 32x24x16 12 Product Usage: is a spinal device that is implanted in the intervertebral body space via an anterior approach to improve stability of the spine while supporting fusion.

FDA Recall Z-2355-2020 · Posted None

Recall StatusOpen, Classified
Product CodeOVD
Date Initiated2019-01-07
Product QuantityA total of 27 lots producing 178 units

Reason for Recall

Due to a manufacturing process issue, it allowed vertebral body bone screws to not seat properly within the spinal device/cage.

Firm Action

On 12/14/2018, the firm notified its supplier via email requesting to visually review its inventory which determined that some screw holes were oval shaped. On 7/23/2019, all inventory was quarantined as part of the removal and was scrapped from inventory. Due to the low risk to patients and low probability, the firm ownly initially begin removal of product from internal locations (main warehouse) for the purpose of inspection and if necessary, repair or destruction. Removal Summary No additional customer or distributor communication took place during the removal.

Root Cause

Process control

Distribution Pattern

US Nationwide distribution including in the states of TX, FL, NC, LA, KY, WI, SD, IL, CA, and KS. O.U.S.: None