Microdot Xtra blood glucose test strips

FDA Recall Z-2357-2019 · Posted None

Recall StatusOpen, Classified
Product CodeNBW
Date Initiated2018-07-26
Product Quantity17,784 (USA)

Reason for Recall

The firm reported that a faulty cutting/punching machine did not meet specifications after being fitted with new blades and did not correctly "notch" BG test strips intended for use in the US. The impacted strips did not fit into the inlet port of the BG meter, therefore resulting in a delay in BG testing and results.

Firm Action

On July 26, 2018, the firm sent an email to it's US distributor that requested a sub-recall.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Illinois

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.