CombiDiagnost R90 Software Version R1.0 and R1.1
FDA Recall Z-2357-2021 · Posted None
Advertisement
Recalling FirmPhilips Medical Systems Gmbh, DMC
Recall StatusOpen, Classified
Product CodeJAA
Date Initiated2020-07-22
Product Quantity322 systems; 66 in US
Reason for Recall
Philips has become aware of a software issue, where the kV and mA values are not locked immediately but change when the collimator shutter position is changed directly after the Lock-In command. This software issue leads to an over or under exposed image.
Firm Action
Philips sent customers an Electronic Product Radiation Correction letter dated July 14, 2021. The letter identified the affected product, problem and actions to be taken.
Philips plans to correct the defect at no cost to customers by implementing a field correction involving the installation of a software update. For questions contact the Customer Care Solutions Center 1-800-722-9377 and reference FCO70900051.
Root Cause
Radiation Control for Health and Safety Act
Distribution Pattern
US Nationwide Distribution