Invacare Pronto Air Personal Transporter (power wheelchair), Invacare Corporation, One Invacare Way, Elyria, Ohio, 44035-4190, (800) 333 -6900.

FDA Recall Z-2358-2015 · Posted 2015-08-10

Recalling FirmInvacare Corporation
Recall StatusTerminated
Product CodeITI
Date Initiated2015-02-02
Date Terminated2015-09-14
Product Quantity1,057 finished devices

Reason for Recall

The batteries used in the Invacare Pronto Air Personal Transporter models power wheelchairs have the possibility to overheat during charging which could potentially result in leaking battery acid and/or swelling battery, loss of use, chemical burns and/or property damage.

Firm Action

Invacare sent an " Urgent Medical Device Recall " letter on April 6, 2015, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. You should immediately contact your provider at the number below so a replacement batteries can be installed. If you have any questions concerning these instructions, please call Invacares recall support line at (877) 413-6008, Monday Friday, 8 a.m. 5 p.m. EST, for assistance. Visit www.invacare.com/recall for more information about this recall.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution - US (nationwide) Distribution including Puerto Rico, and to the states of : AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA and WI., and to the countries of : AUSTRALIA, CANDA, FRANCE, NEW ZEALAND, SOUTH AFRICA, TRINIDAD AND TOBAGO