Biomet Outer Sheath 4.5/5.0,6.5- Orthopedic Manual Surgical Instrument Product Code: 14235
FDA Recall Z-2362-2020 · Posted None
Advertisement
Recalling FirmBiomet
Recall StatusTerminated
Product CodeLXH
Date Initiated2020-05-01
Date Terminated2021-07-28
Product Quantity136 units
Reason for Recall
Device and tray that houses the device did not pass steam sterilization process validation testing
Firm Action
Zimmer Biomet issued letter dated 5/1/20 to Distributors,Risk Managers and Surgeons, states reason for recall, health risk and action to take: Risk Manager Responsibilities:
1. Review this notification and ensure that affected personnel are aware of the contents.
2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility.
3. If the product has been further distributed, provide your customers with the recall notice for hospitals and ensure documentation.
4. Complete Attachment 1 Certificate of Acknowledgement and send to
[email protected]. This form will be returned even if you do not have affected products at your facility.
5. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facilitys documentation.
6. If you have further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 am and 5:00pm EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, your questions may be
emailed to [email protected].
Surgeon Responsibilities:
1. Review this notification for awareness of the contents.
2. There are no specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule
Distributor Responsibilities:
Review this notification and ensure that affected team members are aware of the contents.
2. Immediately locate and quarantine affected product in your inventory.
1. Immediately return all affected product from your distributorship and from affected hospitals within your territory.
Root Cause
Process design
Distribution Pattern
Nationwide Foreign: CHINA, AUSTRALIA, CHILE, COSTA RICA , NETHERLANDS, SWITZERLAND
Official Source