Merge PACS software. The firm name on the label is Merge Healthcare, Inc., Hartland, WI. Intended to create and display two-dimensional and three-dimensional images of anatomy from a series of digitally acquired images.
FDA Recall Z-2363-2016 · Posted None
Advertisement
Recalling FirmMerge Healthcare, Inc.
Recall StatusTerminated
Product CodeLLZ
Date Initiated2016-02-05
Date Terminated2016-10-25
Product Quantity49 sites potentially have the affected versions
Reason for Recall
When measuring a lesion on an unmagnified mammography image then performing the same measurement on an image magnified by the Hologic imager (different image), the measurements are not the same.
Firm Action
The recalling firm's customer support notified customers via phone call beginning 2/5/2016 to inform them of the issue. Customers were informed that a software fix that corrects the issue was available. Customers were asked not to use the measurements on mammo images until upgraded. If customers want an upgrade, then someone would contact them to schedule the upgrade.
Root Cause
Software design
Distribution Pattern
Worldwide Distribution -- USA and Australia.