Invacare Precise Rx Pediatric Flowmeter Model Number: IRCPF16 AW. Packed one per carton. The lnvacare Precise Rx" Pediatric Flowmeter adapts to lnvacare Concentrators for low-flow usage.The gauge has a scale printed on the front side and the back side. The front scale is the primary source for monitoring oxygen flow. The rear scale is an aide for monitoring oxygen flow.

FDA Recall Z-2365-2016 · Posted 2016-08-08

Recalling FirmInvacare Corporation
Recall StatusTerminated
Product CodeNFB
Date Initiated2016-06-21
Date Terminated2017-02-07
Product Quantity395 units

Reason for Recall

The rear gradation scale was screen printed upside down on the flowmeter gauge.

Firm Action

lnvacare Corporation sent an Urgent Medical Device Recall letter dated June 27, 2016, to all affected consignees with return receipt to notify them of the recall. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees with questions were instructed to call 877-413-6008.

Root Cause

Process control

Distribution Pattern

Worldwide Distribution - US to CA, CO, CT, FL, GA, IA, ID, IL,IN,KS,KY,LA,MA,MD,MI,MS,MT,NC,NE,NJ,NM,NY,OH,PA,SC, TN,TX,UT,VA,WA and Internationally to Canada, Germany and New Zealand.