HUDSON RCI, Sheridan Sher-I-BRONCH, Endobronchial Tube (Left sided), Sizes: 28 Fr., 35 Fr., 37 Fr., 39 Fr. and 41 Fr., Rx Only, Sterile. Intended for use in thoracic surgery, bronchospirometry, for the administration of bronchial anesthesia and other uses commonly requiring bronchial intubation.

FDA Recall Z-2367-2015 · Posted 2015-08-20

Recalling FirmTeleflex Medical
Recall StatusTerminated
Product CodeCBI
Date Initiated2015-06-01
Date Terminated2018-04-16
Product Quantity188,195 total units

Reason for Recall

The double swivel connector may crack or separate on the endobronchial tube.

Firm Action

Teleflex Medical sent an Urgent Medical Device Recall Notification letter dated June 1, 2015 to their customers. The letter identified the affected product, problem and actions to be taken. Customers were directed to contact any customers who had received the product included within the scope of this recall. Customers were directed to immediately discontinue use and quarantine any products with the referenced catalog numbers and lot numbers listed in the notification. Additionally, customers were asked to return any affected product together with a completed Recall Acknowledgement Form, for consolidation and return to Teleflex Medical. For questions contact your local sales representative or Customer Service at 1-866-246-6990.

Root Cause

Unknown/Undetermined by firm

Distribution Pattern

Worldwide Distribution - US (nationwide) and the countries of: The Bahamas, Belgium, Argentina, Canada, China, India, Japan, South Korea, Mexico, New Zealand and Vietnam.