Disclaimer: This site aggregates publicly available data from official government sources (FDA, ClinicalTrials.gov, PubMed, SEC EDGAR) for general reference only. It does NOT constitute medical advice, diagnosis, treatment recommendations, or investment advice.

Terminated

Regard SRHS GIC Port A Cath, Sterile, Qty 20/case - Product Usage: The Saline Flush Syringes are intended to be used only for the flushing of indwelling of vascular access devices.

FDA Recall Z-2368-2020 · Posted None

Recall StatusTerminated
Product CodeOFF
Date Initiated2020-05-04
Date Terminated2022-09-23
Product Quantity9,360 units

Reason for Recall

There is a potential for the product to exhibit holes in the packaging, which impacts package integrity and potentially compromises a sterile field.

Firm Action

The firm's supplier mailed an Urgent: Field Action Notification dated 04/21/2020 The notification informs the customers to remove the affected product from inventory and place it in quarantine immediately. The notification included a label template for printing the appropriate warning label. The customers will affix the warning label on all affected kits affected by the notification in inventory. The label should be applied to each affected convenience kit and product case in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. Labels can be printed. The customers should send quantity and end user information request to the contact listed. The kits do NOT need to be returned and may be used. The letter states, an alternate saline flush solution should be used to accompany patient procedures until field action inventory is depleted. The notification states, if any of the affected products identified above were further distributed, to ensure all end-users are appropriately notified of the product recall. Recall Response Forms should be completed and returned as soon as possible even if customers have no current product inventory.

Root Cause

Under Investigation by firm

Distribution Pattern

US Nationwide distribution including in the states of MO, NC.

Official Source

View FDA Recall →

Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.

↑