Xience Sierra TM Everolimus Eluting Coronary Stent System, RX 3.5mm x 33mmStrength: 100 g/cm, Nominal Everolimus Content: 209 g; Lot number 903224A,
FDA Recall Z-2370-2020 · Posted None
Advertisement
Recalling FirmAbbott Vascular
Recall StatusTerminated
Product CodeNIQ
Date Initiated2020-03-25
Date Terminated2021-07-12
Product Quantity13 units
Reason for Recall
Incorrect expiration date
Firm Action
On March 25, 2020, the firm informed their customers that the firm has become aware that products were labeled with the incorrect expiration date. The firm sent an "Urgent Medical Device Recall" to customers. The recalled devices were shipped with a 36 months labeled expiration date, instead of 12 month expiration date.
The firm is instructing customers to :
- Review their inventory and immediately stop using devices from this lot
- Complete and return the attached Effectiveness Check Form
- Return all unused affected product to the firm.
- Share this notification with other relevant personnel in their organization
The firm requested customers to report any adverse reactions or quality problems to Customer Service Department at (800) 227-9902 and to return the "Effectiveness Check Form" to Customer Service Department at (800) 227-9902, scanned copy via E-mail to [email protected]; return a copy of this form with the returned product.
Root Cause
Incorrect or no expiration date
Distribution Pattern
US Nationwide distribution including in the states of NJ, HI, MD, FL, NV, CA, KS and PR. No foreign consignee
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.