BD 60 ml Syringe, Luer-Lok Tip Sterile; Cases : 160 (4 x 40) Shelf Carton: 40 BD; Franklin Lakes, NJ 07417 USA. Made in USA. Syringes for use in aspiration and injection of medications.
FDA Recall Z-2374-2008 · Posted 2008-09-18
Advertisement
Recalling FirmBecton Dickinson & Company
Recall StatusTerminated
Product CodeFMF
Date Initiated2008-07-15
Date Terminated2009-04-16
Product Quantity1246 lots
Reason for Recall
Becton Dickinson identified the presence of open seals during an internal inspection process.
Firm Action
Urgent: Product Recall letters were sent to Distributors and Syringe Customers on July 15, 2008 by UPS second day air. Customers are to immediately review their inventory, determine if they have any affected product, and remove the affected if they have any. Next customers should complete the enclosed Customer Packing List and BD will send a replacement product upon receipt of the returned product. Questions are to be directed to BD Customer Service at 1-888-237-2762, M-F, 8:00 AM to 5:00 PM EST.
Root Cause
Package design/selection
Distribution Pattern
Nationwide Distribution.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.