FormPutty Bone Void Filler 10cc The product is sealed in a foil pouch and then placed in a box following sterilization. Part # 50-07-0100 Is a flowable resorbable porous bone void filler composed of beta tricalcium phosphate and other calcium phosphates.

FDA Recall Z-2375-2008 · Posted 2008-09-18

Recalling FirmTheken Spine LLC
Recall StatusTerminated
Product CodeMQV
Date Initiated2008-07-11
Date Terminated2011-09-20
Product Quantity33 syringes

Reason for Recall

The patient label inside the package has an expiration date of 2008-06 when the actual expiration date is 2010-06.

Firm Action

In a letter dated July 11, 2008, the firm asked their customers to check their inventory for the suspect product. If found they are to return the product for credit or replacement. The firms asks customers to contact them if any units were used for surgery and they will send the correct patient labels. If additional information is needed, customers are to contact Theken Spine Regulatory Affairs or the Quality Department.

Root Cause

Labeling mix-ups

Distribution Pattern

Nationwide Distribution --- including states of FL, OH, TX, and UT.