NORMOFLO Irrigation Fluid Warmer Models H-1100 and H-1129

FDA Recall Z-2376-2021 · Posted None

Recall StatusOpen, Classified
Product CodeLGZ
Date Initiated2021-08-06
Product Quantity913

Reason for Recall

There is potential for harmful levels of aluminum leaching when using lower flow rates with certain solutions and blood products and longer duration of use.

Firm Action

Smiths Medical sent an Urgent Medical Device Recall Notice and attachments dated August 13, 2021 to all affected customers. The letter instructed all customers who purchased affected devices listed in the table on page 1 of the Urgent Medical Device Recall Notice must identify any of the affected products within their possession and refer to the detailed instructions on pages 2 and 3 of the notice. Distributors will be instructed to immediately forward a copy of the Urgent Medical Device Recall Notice and attachments to any customers to whom they have distributed affected product. Distributors will also be instructed to request that customers to who they have distributed affected product complete the Response Form and return it to the distributor. Distributors will indicate their identity as the distributor and document the consignees name and address. Questions regarding this recall notification may be forwarded to [email protected].

Root Cause

Device Design

Distribution Pattern

US Nationwide Distribution and PR OUS distribution in - Austria, Australia, Belgium, Switzerland, Canada, Germany, Spain, Finland, France, United Kingdom, Greece, Hong Kong, Hungary, Ireland, Italy, Kuwait, Luxembourg, Netherlands, New Zealand, Panama, Poland, Portugal, Sweden, Turkey, Venezuela, South Africa