Terminated

Baxter PrismaFlex, REF numbers: a) 106696 (M 60 SET) b) 106697 (M 100 SET) Expansion: c) 107144 (TPE2000 SET) d) 115307 (M 150 SET) e) 115309 (ST 100 SET) f) 955467 (M 150 SET) g) 955468 (ST 100 SET) h) 955595 (M 100 SET) i) 955596 (ST 150 SET) Used with Prismaflex or PrisMax control units for continuous fluid management and renal replacement therapies.

FDA Recall Z-2378-2021 · Posted None

Recall StatusTerminated
Product CodeKDI
Date Initiated2021-07-22
Date Terminated2024-01-17
Product Quantity64275 units

Reason for Recall

Products labeled as sterile were distributed, but may not have been sterilized. " Baxter confirmed Steril Milano observed nonconformances that were not reported to Baxter. Data was manipulated to appear to be conforming

Firm Action

The firm notified its consignees of the recall by letter on 07/22/2021. The letter explained the issue and requested the return of the affected product. Distributors were directed to notified their consignees.

Root Cause

Vendor change control

Distribution Pattern

US Nationwide distribution.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.