MY01 Continuous Compartmental Pressure Monitor, Reference Number MY01-0001
FDA Recall Z-2384-2021 · Posted None
Advertisement
Recalling FirmMY01, INC.
Recall StatusTerminated
Product CodeLXC
Date Initiated2021-06-28
Date Terminated2022-07-25
Product Quantity37 units
Reason for Recall
There is an incorrect version of the needle in the introducer part of devices. The affected Lots were manufactured using needles with a needle trocar geometry which is out of specification and may require a slightly higher push force from the user to penetrate the patient skin.
Firm Action
A Medical Device Advisory Notice dated 6/28/21 was sent to customers.
Clinicians Responsibilities/ Actions required:
1. Review this notification for awareness of the contents
2. Identify and quarantine any unused affected products in your inventory for return
to MY01 Inc. Contact Customer Success team by phone (+1 (855) 292-6901) or
email ([email protected]) to arrange for your replacements.
3. There are no specific patient monitoring instructions related to this voluntary
product removal that are recommended beyond your existing follow up schedule.
4. Complete Attachment 1- Certificate of acknowledgement and email to
[email protected] by 07-Jul-2021.
5. Include a copy of Attachment 1- Certificate of acknowledgement with product
to be returned. Please use the return shipping label (to be provided by MY01
Inc.), please return affected product to:
Regulatory Affairs and Quality Management
MY01 Inc.
400 Maisonneuve Boulevard West,
Suite 700,
Montreal, Quebec,
Canada
H3A 1L4
6. Retain a copy of this document and Attachment 1- Certificate of
acknowledgement with your records.
7. If you have further questions or concerns after reviewing this notice, please call
the customer service team at 1-855-292-6901. Alternatively, your questions may
be emailed to: [email protected]
Under 21 CFR 803, manufacturers are also required to report any serious injuries where
a product has contributed or may have contributed to the event. Please keep MY01 Inc.
informed of any adverse events associated with this by emailing [email protected] or call
+1(855)-292-6901.
Root Cause
Under Investigation by firm
Distribution Pattern
US Nationwide distribution in the states of MD, OH, CA, NY.