Recall StatusTerminated
Date Initiated2010-08-02
Date Terminated2012-07-02
Product Quantity3106 units
Reason for Recall
The recall was initiated after Beckman Coulter confirmed reports that black / brown particulates were observed in some bottles of the Creatinine Alkaline Buffer (Buffer lot M911427 and M911518) in modular Creatinine (CREm) reagent kit lots M911428 and M911519. Beckman Coulter have had reports of erroneous results when using the lots of Creatinine Alkaline Buffer listed above. Single
creatinine d
Firm Action
The firm, Beckman Coulter, sent a "URGENT: PRODUCT CORRECTIVE ACTION" letter dated August 3, 2010, to all customers. The letter describes the product, problem, and action to be taken by customer. The customers were instructed to:
(1) Please discontinue use of Creatinine (CREm) reagent kit lots M911428 and M911519 and discard appropriately.
(2) Please contact Beckman Coulter at (800) 526-3821 in the United States or (800) 463-7828 in Canada to discuss
CREm kit M911428 and M911519 replacement options. Outside the United States and Canada, please contact
your local Beckman Coulter Representative.
(3) Please complete and return the enclosed response form within 10 days so that we may maintain our records.
(4) Please share this information with their laboratory staff and retain this notification as part of their Quality System
documentation. If you have forwarded the affected products listed above to another laboratory, please provide a copy of this letter to them.
If customers have any questions regarding this Product Corrective Action letter, they were instructed to please contact Beckman Coulter's Customer Support Center (Call Center) on the Beckman Coulter website at www.BeckmanCoulter.com/Customer Support/Support
Requests, or call (800) 854-3633 in the United States and Canada.
Distribution Pattern
Worldwide Distribution - USA including the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KY, KS, LA, MA, ME, MD, MI, MN, MO, MS, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, TN, TX, VA, WA, WI, WV, and WY and country of Canada
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.