Hospira brand Symbiq Two-Channel Infuser, LIST No. 16027 Symbiq Two-Channel Infuser, List Number Configurations: 16027-04-01/03/04, 51/52, 53/54, 79/80,81/82,83/84,87/88 and 89/90 16026-27-89/90, 16027-27-89/90, 16027-13-25/26,53/54, Distributed by: Hospira, Inc. Morgan Hill, CA

FDA Recall Z-2391-2010 · Posted 2010-09-09

Recalling FirmHospira Inc
Recall StatusTerminated
Product CodeFRN
Date Initiated2010-02-22
Date Terminated2011-04-01

Reason for Recall

The clamp may not secure to the IV pole and the pump may slip or fall off. The pump may fall on a person or pull the tubing out of the IV container.

Firm Action

Hospira issued an "Urgent: Device Field Correction" Letter dated February 22, 2010 to all consignees, identifying the reason for the recall and affected product. Customers were informed that they will be contacted by a Hospira representative to replace the affected pole clamps. Consignees can contact Hospira Technical Support Operations at 1-800-241-4002.

Root Cause

Device Design

Distribution Pattern

Worldwide Distribution: USA (all states and Puerto Rico) and the countries of Australia, Canada, Colombia, Costa Rica, Malaysia, Mexico, Singapore. and the UAE.