ES0102, AEM Disposable Curved Scissors, 3/4", 35cm, Sterile R, Encision Inc., Boulder, CO 80301 Cutting tissue during electrosurgical procedures
FDA Recall Z-2396-2011 · Posted 2011-05-31
Advertisement
Recalling FirmEncision, Inc.
Recall StatusTerminated
Product CodeGEI
Date Initiated2011-04-07
Date Terminated2011-08-09
Product Quantity952 Ten-Count Boxes
Reason for Recall
Sterility may be compromised due to holes in packaging.
Firm Action
Encision, Inc. sent an Important Safety Notice letter dated April 7, 2011, to all affected customers. The letter identified the product, the problem, and the action to be taken by the customer. Customers were instructed to either carefully inspect any remaining pouches for readily visible holes in the Tyvek pouch, or return the affected lots to Encision for inspection.
Customers were also asked to fill in and fax back the attached Confirmation to 303-444-2693.
Root Cause
Package design/selection
Distribution Pattern
Worldwide Distribution - USA (nationwide) and the countries of Australia and New Zealand
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.