Ingenia Elition S Magnetic Resonance (MR) system, Product Numbers 781357 and 782106
FDA Recall Z-2396-2023 · Posted 2023-08-11
Advertisement
Recalling FirmPhilips North America
Recall StatusOpen, Classified
Product CodeLNH
Date Initiated2023-06-12
Product Quantity640 total
Reason for Recall
The system operator is able to bypass the SmokeDetector Interlock system locking mechanism after a smoke detection by power cycling the system.
Firm Action
An URGENT Medical Device Correction Notice dated 6/5/23 was sent to customers.
Actions that should be taken by the customer / user in order to prevent risks for patients or users:
4.1. Philips has installed a SmokeDetector Interlock feature on all the systems listed in Appendix A.
Post this notice near the affected MR system(s) for ease of reference.
Circulate this notice to all users of this device so that they are aware of the product issue
and associated hazard/harm until this issue has been resolved.
Please complete and return the attached updated customer response form to Philips
promptly and no later than 30 days from receipt of this letter. Philips is requesting that
even if you acknowledged a previous version of this letter that you also respond with an
acknowledgment to this updated letter.
4.2. The SmokeDetector Interlock feature is integrated with the MR System and powers off the relevant part of the system (Gradient Amplifier) when smoke is detected in order to prevent further development of smoke and/or fire. If the SmokeDetector Interlock has detected smoke and has powered off the Gradient Amplifier:
Remove patient from the system according to the instructions for use because power
to the tabletop will still be active
Scanning is disabled until Philips service has checked the system. Do not attempt to
continue scanning.
Immediately inform Philips Service
DO NOT attempt to restart or perform a power cycle of the system
4.3. If a user experiences smoke, fire, or severe image quality issues (image artifacts caused by electrical disturbances):
Immediately stop scanning and evacuate the patient from the Examination Room.
Check Examination Room for a developing fire.
If a developing fire is sensed, adhere to established fire emergency procedures, which
may include switching off power to the complete system and/or removing the magnet
field by using the Emergency Magnet Off button
Scanning is disabled until Philips service has checked t
Root Cause
Under Investigation by firm
Distribution Pattern
US Nationwide. Global Distribution.
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
K213516Ingenia 3.0T, Ingenia 3.0T CX, Ingenia Elition and MR 7700 with Distributed Multi NudcleiPhilips Medical Systems Nederland B.V. · 2022-03-03K193215Achieva, Intera, Ingenia, Ingneia CX, Ingenia Elition, and Ingenia Ambition MR SystemsPhilips Medical Systems Nederland B.V. · 2020-04-10K183063Ingenia, Ingenia CX, Ingenia Elition, and Ingenia Ambition MR SystemsPhilips Medical Systems Nederland B.V. · 2019-02-14K173451Ingenia Elition S and Ingenia Elition XPhilips Medical Systems Nederland B.V. · 2018-03-20K173356Ingenia Elition S and Ingenia Elition X R5.4Philips Medical Systems Nederland B.V. · 2018-03-06