Hintermann Series Talar Implant, Left, Size 2 REF 302112
FDA Recall Z-2399-2026 · Posted 2026-06-10
Advertisement
Recalling FirmDT MedTech, LLC
Recall StatusOpen, Classified
Product CodeNTG
Date Initiated2026-03-24
Product Quantity67 units
Reason for Recall
Through customer complaints, the ankle replacement system was found to contain incorrect devices: Left implant packaging and labeling bears a right implant device.
Firm Action
On March 24, 2026, DT MedTech issued a Urgent Medical Device Recall Notification to affected consignees. DT MedTech requested that the firm take the following actions:
1. Return of all available stock of Hintermann Series Talar Implants (Part/REF number 302112, Lot number AACAA)
2. Please provide confirmation that all available stock of the part detailed below has been returned to Vilex LLC at 4516 S 700 E Suite 100A Salt Lake City, Utah 84107 USA.
3. For any questions or concerns, contact Vilex Quality at [email protected] or 800-521-5002.
Root Cause
Labeling Change Control
Distribution Pattern
Worldwide - US Nationwide distribution in the states of CA, AR, AZ, ID and the countries of PER, KOR.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.