PLAC Test ELISA kit, catalog number 90106, enzyme immunoassay for the quantitative determination of Lp-PLA2 in human plasma and serum, kit containing coated microwell stripplate and reagents (12 strips, 1 set calibrators 1 to 6, 0.25 ml each, 20x wash buffer, 50 mL, conjugate 23 mL, TMB, 11 mL, Stop solution, 11 mL

FDA Recall Z-2400-2008 · Posted 2008-09-26

Recalling FirmDiaDexus, Inc
Recall StatusTerminated
Product CodeNOE
Date Initiated2008-07-18
Date Terminated2009-12-22
Product Quantity65 (total)

Reason for Recall

Product may give Lp-PLA2 values at up to 22% lower than accurate values.

Firm Action

All customers received communication of product correction dated July 18. 2008, via Federal Express, which included a cover letter, technical bulletin, specimen handling best practices cards, and revised package inserts. E-mail communication was also sent to all but seven customers. Contact diaDexus Technical Support at 1-877-752-2837 for assistance.

Root Cause

Finished device change control

Distribution Pattern

US, OUS: UK and Germany.

Related FDA 510(k) Records

Devices referenced by this recall via K-numbers.