Stratus CS Acute Care Troponin I TestPak ( Troponin I assay)
FDA Recall Z-2400-2010 · Posted 2010-09-11
Advertisement
Recalling FirmSiemens Healthcare Diagnostics, Inc.
Recall StatusTerminated
Product CodeMMI
Date Initiated2010-04-13
Date Terminated2010-10-25
Product Quantity7500 cartons
Reason for Recall
False elevated Stratus CS CTNI results
Firm Action
Recalling firm issued " Urgent Field Safety Notice" dated April 2010, informing users of the problem and to discuss with laboratory directors regarding the need to review previous test results, conduct patient follow up, and/or repeat testing for tests conducted on or after Nov 1, 2009.
Root Cause
Process control
Distribution Pattern
Nationwide Distribution