Invacare Foot Section for Semi & Full Electric Beds, Junction Boxes, Product Usage: The Invacare Bariatric home care bed is intended to provide a rest surface with adjustable height from floor and adjustable surface contours.
FDA Recall Z-2401-2016 · Posted None
Advertisement
Recalling FirmInvacare Corporation
Recall StatusTerminated
Product CodeFNL
Date Initiated2016-03-15
Date Terminated2017-09-15
Product Quantity3,050 units
Reason for Recall
The firm recently identified a quality issue with the Junction box used in the Invacare Bariatric Bed. They discovered that the Junction Box may omit sparks which could result in a fire in the presence of flammable materials.
Firm Action
Invacare sent an Urgent Medical Device Recall letter dated May 12, 2016 to affected customers. The letter identified the affected product, problem and actions to be taken. For questions call Invacares recall support line at (877) 413-6008, Monday Friday, 8 a.m. 5 p.m. EST, for assistance.
Root Cause
Nonconforming Material/Component
Distribution Pattern
US Nationwide Distribution including DC and PR. Internationally Canada.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.