Terminated

SOMATOM go.All; Models #11061630 Product Usage: This computed tomography system is intended to generate and process cross-sectional images of patients by computer reconstruction of x-ray transmission data

FDA Recall Z-2402-2019 · Posted None

Recall StatusTerminated
Product CodeJAK
Date Initiated2019-07-31
Date Terminated2020-12-02
Product Quantity10 units

Reason for Recall

The injector holder could potentially fall off together with the injector and the storage box.

Firm Action

1. Take special care when rotating the injector and storage box horizontally 2. Your Siemens Healthineers service engineer will perform an on-site check to evaluate the stability of the injector arm. 3. If there is any problem with the injector arm, your Siemens Healthineers service engineer will replace the injector arm with an improved intermediate version free of charge. 4. Please ensure all users of the affected products are aware. 5. Retain this notification in your product related records and place a copy in the medical device's instructions for use. 6. If you have distributed this product, identify your customers and notify them. If you wish to request an earlier appointment or have additional questions, feel free to contact service organization at 1-800-888-7436 .

Root Cause

Nonconforming Material/Component

Distribution Pattern

US Nationwide Distribution in the states FL KS WI NE MS OH TX NY LA IN IL OK MI GA NJ AK CT AL MD