Capsule Neuron UMPC is a component in the DataCaptor" Connectivity System. Catalog Number: DC-NU-MPC Capsule Neuron" UMPC is a bedside device that enables automatic collection of vital signs data that can local cache data and connect to the hospital network through a wired Ethernet connection or via a standard 802.11 b/g/n wireless network, available in a fixed or mobile mounting option, running either DataCaptor continuous or periodic connectivity software The DataCaptor" System is indicated for use in data collection and clinical information management either directly or through networks with independent bedside devices. DataCaptor" is not intended for monitoring purposes, nor is the software intended to control any of the clinical devices (independent bedside devices / information systems) it is connected to
FDA Recall Z-2403-2011 · Posted 2011-06-01
Advertisement
Recalling FirmCapsule Tech Inc.
Recall StatusTerminated
Product CodeMWI
Date Initiated2010-08-25
Date Terminated2012-03-20
Product Quantity3,144 Capsule Neuron UMPC units
Reason for Recall
Software:
Data cached on the Neuron due to server unavailability is not being delivered to the server after server availability is restored.
If a compact flash failure occurs, the file system may be corrupted and result in system hangs, lock-ups or crashes.
Batteries can discharge completely in a Capsule Neuron" UMPC even though it is connected to power source
Firm Action
Capsule Tech initiated three corrective action notifications for the Neuron" UMPC units three failure modes as follows:
a) Data cached on the Capsule Neuron" UMPC due to server unavailability is not being delivered to the server after server availability is restored:
i) A customer advisory was provided to a subset of consignees, along with DataCaptor software version 6.6.5 on December 1, 2010.
ii) A customer advisory was provided to the remaining consignees, along with DataCaptor software version 6.6.7 on January 27, 2011.
If a compact flash failure occurs, the file system may be corrupted and result in system hangs, lock-ups or crashes:
i) A customer advisory was provided to all consignees on March 23, 2011, via phone call, email and FedEx;.The correction will be provided as a software update. A formal recall notification letter will be provided to all consignees alerting them to the availability of the correction once it is released, via phone call, email and FedEx.
c) Batteries can discharge completely in a Capsule Neuron" UMPC even though it is connected to power. Note: the Capsule Neuron" UMPC operates normally when connected to power.
i) A customer advisory was provided to all consignees on April 23, 2010, via FedEx.
For questions regarding this recall contact Capsule technical support team at
(800) 260-9537 or e-mail [email protected]..
Root Cause
Software design
Distribution Pattern
Worldwide Distribution - USA (nationwide) and the country of Canada
Related FDA 510(k) Records
Devices referenced by this recall via K-numbers.
Official Source
View FDA Recall →Source: FDA Enforcement Report / Recall database. Always confirm against the official FDA record.