Terminated

Medicine Cup, 1 OZ. Liquid medication dispenser to issue a measured amount of liquid medication.

FDA Recall Z-2404-2016 · Posted None

Recall StatusTerminated
Product CodeKYW
Date Initiated2016-05-11
Date Terminated2019-06-03
Product Quantity903 Cases and 4,090 Sleeves (8,605,000 Medicine Cups)

Reason for Recall

There is 5 mL printed next to the increment on the medicine cups instead of 10 mL. The cup has the correct mL markings for the other gradations, but sequentially, the markings are listed as 2.5 mL, 5 mL, 7.5 mL, and 5 mL.

Firm Action

Medline Industries sent a an URGENT RECALL letter dated 5/11/2016 to all affected consignees via certified mail. The URGENT RECALL letter described the reason for the recall and the actions to take.

Root Cause

Process control

Distribution Pattern

Continental U.S.