MultiDiagnost-Eleva
FDA Recall Z-2404-2023 · Posted 2023-09-07
Advertisement
Recalling FirmPhilips Medical Systems Nederland B.V.
Recall StatusOpen, Classified
Product CodeKPR
Date Initiated2023-07-19
Product Quantity5432 systems in total
Reason for Recall
A foot switch pedal may get stuck in the active position when the user releases the pedal,
resulting in the emission of unintended radiation
Firm Action
Philips will send a customer notification to the existing users of the affected systems to alert them of the issue and provide actions that should be taken by the customer/user in order to prevent risks to patients. Philips will contact customers to schedule a visit to inspect the foot switch and provide an addendum to the Instructions for Use (IFU) of the system detailing the appropriate cleaning and maintenance instructions. (Philips reference 2023-IGT-BST-011).
Root Cause
Radiation Control for Health and Safety Act
Distribution Pattern
US Nationwide Distribution