Precise Digital Accelerator Delivery of radiation to defined target volumes

FDA Recall Z-2405-2012 · Posted 2012-09-19

Recalling FirmElekta, Inc.
Recall StatusTerminated
Product CodeIYE
Date Initiated2012-08-06
Date Terminated2014-10-08
Product Quantity110

Reason for Recall

There have been a small number of instances in hospitals where the joint between the drive gear and the gantry base drive wheel has become loose, leading to fatigue failure of some of the securing bolts.

Firm Action

Elekta sent an Important Field Safety Notice dated August, 2012 to inform users that inspection and possibly bolt replacement needs to be carried out. Users should return the signed acknowledgement to Elketa, and an inspection should be carried out under FCO20002103053. For questions customers should call +44(0)1293 654200. For questions regarding this recall call 770-300-9725.

Root Cause

Nonconforming Material/Component

Distribution Pattern

Nationwide Distribution including AL, AZ, CA, CO, CT, FL, GA, IL, IA, MD, MA, MI, MN, NE, NV, NJ, NM, NY, ND, OH, OK, OR, PA, VA, SC, TN TX, WA, WV, WI and Puerto