RenLane Renal Denervation Catheter; Cat No.: D135601. Used in adults patients (>18 years) with drug resistant hypertension to denervate the renal arteries to reduce blood pressure.
FDA Recall Z-2405-2015 · Posted 2015-08-18
Advertisement
Recalling FirmCordis Corporation
Recall StatusTerminated
Product CodeMPB
Date Initiated2014-04-15
Date Terminated2015-08-26
Product Quantity98 units
Reason for Recall
Potential Damage may occur to the helical tip of the RENLANE Renal Denervation Catheter, if the hemostasis valve of the arterial access device is not fully open before insertion or withdrawal of the catheter.
Firm Action
Cordis sent an Urgent Field Safety Notice dated April 15, 2014 providing additional labeling to their consignees in Germany only, with a follow-up letter on August 4, 2014. Cordis explained When deploying the catheter via a compatible arterial access device, if a Tuohy-Borst type adjustable hemostasis valve is used, ensure the valve is fully open before insertion or withdrawal of the catheter. This is especially important, since the helical tip at the distal end of the RENLANE Renal Denervation Catheter has a larger diameter compared to the proximal catheter shaft. Although the event was not a Removal, all lots involved have expired. The product is no longer commercialized, and was never commercialized beyond the 10 customers in Germany.
Root Cause
Labeling Change Control
Distribution Pattern
Distributed in Germany only.