Lase DiscectomLase Discectomy Device Kit, Models 1100-002 and 1100-002 SHORTy Device Kit, Models 1100-002

FDA Recall Z-2405-2021 · Posted None

Recalling FirmClarus Medical, Llc
Recall StatusTerminated
Product CodeGEX
Date Initiated2021-07-27
Date Terminated2022-09-15
Product Quantity858 units

Reason for Recall

The sterile barrier seal may be compromised.

Firm Action

The affected Domestic customers, Domestic/International Distributors will be notified with a Field Safety Notice. Domestic/International distributors will also include a request for a list of the Distributor s customers.

Root Cause

Packaging process control

Distribution Pattern

Distribution in US - 18 accounts 1 foreign account