Utak NR Trace Elements Serum Control, 5 ml, REF 66816. The responsible firm on the label is UTAK LABORATORIES, INC., Valencia, CA. Control for in vitro diagnostics.

FDA Recall Z-2407-2020 · Posted None

Recalling FirmUtak Laboratories Inc
Recall StatusTerminated
Product CodeJJY
Date Initiated2020-03-20
Date Terminated2020-07-22
Product Quantity1,395 vials

Reason for Recall

The product has an incorrect reference value for cadmium and an incorrect expected range.

Firm Action

The recalling firm issued a recall letter dated 3/19/2020 via email on 3/20/2020 requesting customers to destroy the incorrect IFU and replace it with the revised IFU.

Root Cause

Labeling Change Control

Distribution Pattern

Worldwide Distribution. US nationwide, Canada, Australia, France, Germany, Israel, Italy, New Zealand, Philippines, and United Kingdom.